published 24 Jul 2025 · updated 27 Mar 2026
Medtronic MyCareLink Patient Monitor Empty Password Vulnerability
6.8MediumCVSS 6.8EPSS 0%
No exploitation in the wild has been reported. EPSS puts the chance of exploitation in the next 30 days at 0%. Fixed in June 25, 2025.
Medtronic MyCareLink Patient Monitor has a built-in user account with an empty password, which allows an attacker with physical access to log in with no password and access modify system functionality. This issue affects MyCareLink Patient Monitor models 24950 and 24952: before June 25, 2025
Affected
Vendors, products and packages with version ranges
| Vendor | Product | Versions | Fixed |
|---|---|---|---|
| Medtronic | <June 25, 2025, <June 25, 2025 | June 25, 2025, June 25, 2025 |
Timeline
Every event on this vulnerability
Embed
A live severity badge for an advisory, README or ticket