CVE-2025-64298

published 2 Dec 2025 · updated 9 Dec 2025

Mirion Medical EC2 Software NMIS BioDose Incorrect Permission Assignment for Critical Resource

8.4HighCVSS 8.4EPSS 0%

No exploitation in the wild has been reported. EPSS puts the chance of exploitation in the next 30 days at 0%. Fixed in 23.0.

NMIS/BioDose V22.02 and previous version installations where the embedded Microsoft SQLServer Express is used are exposed in the Windows share accessed by clients in networked installs. By default, this directory has insecure directory paths that allow access to the SQL Server database and configuration files, which can contain sensitive data.

Affected

Vendors, products and packages with version ranges

VendorProductVersionsFixed
Mirion Medical<23.023.0

Timeline

Every event on this vulnerability

Embed

A live severity badge for an advisory, README or ticket

CVE-2025-64298 severity on Zeptric[![CVE-2025-64298](https://zeptric.io/badge/cve/CVE-2025-64298.svg)](https://zeptric.io/cve/CVE-2025-64298)